Abstract

The MicroScan system was compared with agar dilution screen plates for the detection of high-level aminoglycoside resistance in 182 enterococcal isolates. Both the frozen Gram-Positive Combo Type 2 and the freeze-dried Type 5 panels were evaluated. The specificity of both panels for the detection of streptomycin and gentamicin resistance was 100%. However, the sensitivities for the detection of gentamicin and streptomycin resistance were 84 and 31%, respectively, for the Type 2 panels and 90 and 41%, respectively, for the Type 5 panels. The sensitivities of these panels for the detection of enterococcal high-level aminoglycoside resistance are inadequate for routine use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.