Abstract

A new, commercially available latex agglutination test (SeroSTAT Staph; Scott Laboratories, Inc., Fiskeville, R.I.) was compared with the tube coagulase and slide coagulase tests as means for identifying Staphylococcus aureus. Of 160 clinical isolates of S. aureus, 159 (99.4%) yielded positive results with the latex agglutination test. Negative latex agglutination test results were obtained with 266 of 267 clinical isolates of Micrococcus spp. and staphylococcal species other than S. aureus (99.6%). The latex agglutination test was found to be a rapid, technically nondemanding method for identifying S. aureus. It was as accurate as the tube coagulase test and more accurate than the slide coagulase test.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.