Abstract

The full-energy peak (photopeak) efficiency and peak-to-total ratio of a closed-end coaxial high-purity germanium detector were calculated and compared with the experimental measurements in the energy range between 60 and 1863 keV. The attenuation of photons by the source itself (self-absorption), the source container, and the detector end cap materials is also included. Many experimental readings have been taken for the source-detector system used in this investigation. By comparison, the calculated and experimental full-energy peak efficiency and peak-to-total ratio values are in good agreement; the overall percentage error is <3%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.