Abstract

The aims of this study were to elucidate to what extent storage and repeated freezing and thawing influenced the concentration of creatinine in urine samples and to evaluate the method for determination of creatinine in urine. The creatinine method was based on the well‐known Jaffe's reaction and measured on a COBAS Mira autoanalyser from Roche. The main findings were that samples for analysis of creatinine should be kept at a temperature of −20°C or lower and frozen and thawed only once. The limit of detection, determined as 3×SD of 20 determinations of a sample at a low concentration (6.1 mmol/L), was 0.3 mmol/L, and the recovery of a certified reference material was 97%. The relative precision at 3.15 mmol/L was 2.3%. It was concluded that the method is appropriate for measurement of urinary creatinine.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.