Abstract

The properties of the electron density phantom CIRS M62 were studied. Physical and electron densities were calculated for all tissue substitutes, and compared to the values given in the data sheet and to the recommendations of the ICRU Report 44. The phantom was scanned in standard-CT and spiral-CT mode with different acquisition parameters. Especially manufactured tissue substitutes for dense bone were analysed for artifacts. Hounsfield values of all tissues substitutes were determined with the treatment planning system CadPlan®. Calibration curves were calculated and compared to a stoichiometric calibration for real tissue. There was good agreement between the values calculated and the values given by the manufacturer. The tissue substitutes were only approximately tissue equivalent. The tissue substitutes for dense bone showed small artifacts. The calculated calibration curves were in good agreement with the stoichiometric calibration curve.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.