Abstract
It is desirable to quickly and accurately determine the pH of reduced-moisture solid food systems. Several techniques of pH determination were evaluated on such systems. Methods involving the addition of water yielded pH values higher than actually existed based on pH-dependent chemical reactions. the Ultra M® micro electrode had problems with reproducibility and stability in the reduced-moisture systems as well as the buffer solutions used for calibration. the Horiba Cardy pH meter and the surface pH electrode used directly on the surface of the reduced-moisture solid system yielded pH values similar to those determined using pH-dependent chemical reaction markers. the ability to measure the actual pH of reduced-moisture solid systems would enable better understanding of potential modifications in product formulation needed to retard microbial growth and limit chemical reactions responsible for loss of shelf-life.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.