Abstract
The goal of radiation therapy is twofold: maximize the possibility of destroy malignant cells while minimizing the damage to healthy tissue. The introduction of intensity modulated radiation therapy (IMRT) technique has brought improvements in this goal. Multi-leaf collimator (MLC) is a useful tool for IMRT. However, the use of MLC is not necessarily mandatory. The Panther Treatment Planning System version 4.6, Prowess Inc., enables the implementation of this technique for accelerator without MLC (the socalled Jaws-Only IMRT technique). This study aims to evaluate the results of application of Jaws-only IMRT technique for nasopharyngeal carcinoma patients at Dong Nai general hospital. Twenty five patients were randomly selected for this study. For each patient, two plans were generated: 3D-CRT (Three-Dimensional Radiation Treatment) and JO-IMRT. The dose distributions, dose-volume histograms (DVH), conformity indexes (COIN), homogeneity indexes (HI) were used to compare between these two plans and find out the best plan. Pretreatment verifications were performed for all patients' plans using ion chamber (Farmer Type Chamber FC65-P, IBA), detector array (MapCHECK2, Sun Nuclear Corporation and Octavius 4D 1500, PTW). The average deviation between measurement and calculation for point dose was 2.3±1.1 %, within limit dose constraint. For detector array measurements, the gamma index with 3 % dose difference and 3 mm was higher than 95 %. The results showed that the JO-IMRT technique had generated better dose distribution in the target volume and reduced dose to healthy tissues compared to 3D-CRT.
Highlights
The dose maximize the possibility of destroy malignant distributions, dose-volume histograms (DVH), cells while minimizing the damage to healthy conformity indexes (COIN), homogeneity indexes tissue
Multi-leaf collimator verifications were performed for all patients' (MLC) is a useful tool for Intensity modulated radiation therapy (IMRT)
For aims to evaluate the results of application of detector array measurements, the gamma index Jaws-only IMRT technique for nasopharyngeal with 3 % dose difference and 3 mm was higher carcinoma patients at Dong Nai general hospital. than 95 %
Summary
Xạ trị điều biến cường độ (IntensityModulated Radiation Therapy, IMRT) là một kỹ thuật tiên tiến trong xạ trị, bắt đầu phổ biến từ 1994 [1, 2]. Mục tiêu của bài báo này là đánh giá kỹ thuật IMRT cho ung thư vòm với hệ máy gia tốc không có MLC tại Bệnh viện Đa khoa Đồng Nai. Kỹ thuật này được gọi là Jaws Only-IMRT (JOIMRT), được thực hiện dựa trên hệ thống phần mềm lập kế hoạch xạ trị Prowess Panther 4.6 (nhà cung cấp Prowess, Mỹ) [3]. Mục đích của nghiên cứu này là đánh giá khả năng áp dụng kỹ thuật xạ trị IMRT bằng phần mềm Prowess Panther cho bệnh nhân ung. Thông thường việc đánh giá một kế hoạch chỉ dựa trên sự phân bố liều của các đường đồng liều trên từng lát cắt CT (phải bao sát khối u) và trên biểu đồ liều lượng theo thể tích DVH (liều tới khối u đúng liều chỉ định và các cơ quan lành nằm trong giới hạn cho phép chấp nhận được) [16]. (2) và sau đó được vẽ dưới hạng biểu đồ hộp bằng phần mềm Origin 8.0
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have