Abstract

BackgroundThe term posthospital syndrome has been used to describe the condition in which older patients are transiently frail after hospitalization and have a high chance of readmission. Since low activity and poor sleep during hospital stay may contribute to posthospital syndrome, the continuous monitoring of such parameters by using affordable wearables may help to reduce the prevalence of this syndrome. Although there have been systematic reviews of wearables for physical activity monitoring in hospital settings, there are limited data on the use of wearables for measuring other health variables in hospitalized patients.ObjectiveThis systematic review aimed to evaluate the validity and utility of wearable devices for monitoring hospitalized patients.MethodsThis review involved a comprehensive search of 7 databases and included articles that met the following criteria: inpatients must be aged >18 years, the wearable devices studied in the articles must be used to continuously monitor patients, and wearables should monitor biomarkers other than solely physical activity (ie, heart rate, respiratory rate, blood pressure, etc). Only English-language studies were included. From each study, we extracted basic demographic information along with the characteristics of the intervention. We assessed the risk of bias for studies that validated their wearable readings by using a modification of the Consensus-Based Standards for the Selection of Health Status Measurement Instruments.ResultsOf the 2012 articles that were screened, 14 studies met the selection criteria. All included articles were observational in design. In total, 9 different commercial wearables for various body locations were examined in this review. The devices collectively measured 7 different health parameters across all studies (heart rate, sleep duration, respiratory rate, oxygen saturation, skin temperature, blood pressure, and fall risk). Only 6 studies validated their results against a reference device or standard. There was a considerable risk of bias in these studies due to the low number of patients in most of the studies (4/6, 67%). Many studies that validated their results found that certain variables were inaccurate and had wide limits of agreement. Heart rate and sleep were the parameters with the most evidence for being valid for in-hospital monitoring. Overall, the mean patient completion rate across all 14 studies was >90%.ConclusionsThe included studies suggested that wearable devices show promise for monitoring the heart rate and sleep of patients in hospitals. Many devices were not validated in inpatient settings, and the readings from most of the devices that were validated in such settings had wide limits of agreement when compared to gold standards. Even some medical-grade devices were found to perform poorly in inpatient settings. Further research is needed to determine the accuracy of hospitalized patients’ digital biomarker readings and eventually determine whether these wearable devices improve health outcomes.

Highlights

  • BackgroundMost physiologic parameters, such as vital signs or activity, are routinely monitored a few times each day in hospital ward settings [1]

  • The included studies suggested that wearable devices show promise for monitoring the heart rate and sleep of patients in hospitals

  • Many devices were not validated in inpatient settings, and the readings from most of the devices that were validated in such settings had wide limits of agreement when compared to gold standards

Read more

Summary

Introduction

BackgroundMost physiologic parameters, such as vital signs or activity, are routinely monitored a few times each day in hospital ward settings [1]. Since subtle changes in vital signs are often present 8 to 24 hours before a life-threatening event, such as intensive care unit admission or cardiac arrest, vital sign surveillance has the potential to detect clinical deterioration at an earlier phase, thereby permitting clinicians to make corrective interventions [4,5,6,7]. This includes identifying patients with poorly controlled pain and recognizing arrhythmias. There have been systematic reviews of wearables for physical activity monitoring in hospital settings, there are limited data on the use of wearables for measuring other health variables in hospitalized patients

Methods
Results
Discussion
Conclusion