Abstract
Recent reports have shown both poly (methyl methacrylate) (PMMA) and silica optical fibres to be ionization quenching free, making them possibly very useful dosimeters for proton beams. In this study, the response from PMMA and silica optical fibres to therapeutic proton beams are evaluated. The light output was recorded from both optical fibres, exposed to varying dose-rates of 0.5 Gy/min to 20 Gy/min from a 235 MeV isochronous cyclotron. The PMMA optical fibre was observed to have a linear dose-rate response, and a constant light emission for a constant dose-rate exposure. However, in the case of the silica optical fibres, the light output was observed to increase during a constant dose-rate exposure. If uncorrected, this accumulated dose sensitivity observed in the silica optical fibres can result in erroneous measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.