Abstract
The evaluation of platelet activation of medical devices using in cardiovascular system is very meaningful. Currently, it is mainly based on the ISO10993-4 international standard. However, the methods given in the standard are originally designed for non-degradable materials, the applicability, the operability, and the convenience to degradable materials of the methods needs to be carefully studied. In this study, the platelet activation by 3 typical degradable materials (collagen, polylactic acid and hydroxyapatite) were evaluated by three widely used molecular markers CD62P, CD63, CD40L and the three molecular markers PF4, β-TG and TXB2 mentioned in the ISO10993-4 standard. The variations of the six markers in a simulated degradation process of the degradable materials were compared. It was found that the degree of platelet activation changed with the degradation and was strongly relative with the surface physicochemical properties. For example, when the surface roughness and contact angle of the materials change, the degree of platelet activation also changes. These six platelet activation molecular markers can to be the promising key for the assessing of platelet function in degradable medical devices which is instructive for the quality control and the development of new degradable medical devices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.