Abstract

The performance of SD Bioline rapid antigen test (RAT) was evaluated using real-time reverse transcription polymerase chain reaction (rRT-PCR) as gold standard. A total of 718 nasal swabs, including 102 rRT-PCR positive and 616 rRT-PCR negative swabs, were tested. RAT demonstrates a sensitivity of 29·4% with a specificity of 100%. The positivity rate of RAT was highly associated with lower cycle threshold (Ct) values (P < 0·0001). The excellent specificity of the RAT allowed for the rapid identification of influenza cases. However, negative results should be verified by rRT-PCR test because of limitations observed in sensitivity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.