Abstract

HyperArc is a preferred technique for treating brain metastases, employing a single isocenter for multiple lesions. Geometrical isocentricity in the TrueBeam linear accelerator with HyperArc is crucial. We evaluated machine performance checks (MPCs) as an alternative to the Winston-Lutz (WL) test to verify the treatment isocenter. Between January and July 2023, we assessed 53 data points using MPC and Winston-Lutz tests. The isocenter size obtained from the MPC and its sum, including the rotation-induced couch shift, were compared with the maximum total delta value from the Winston-Lutz test. The maximum total delta was 0.68 ± 0.10 mm, while the isocenter size was 0.28 ± 0.02 mm. The sum of the isocenter size and rotation-induced couch shift measured by MPC was 0.61 ± 0.03 mm. During the Winston-Lutz test (without couch rotation), the maximum total delta value was 0.56 ± 0.13 mm. A t-test analysis revealed a significant difference in the isocenter size averages between the Winston-Lutz and MPC outcomes, whereas the Pearson's correlation coefficient yielded no correlation. Our study highlights the necessity for separate MPC and Winston-Lutz tests for isocenter verification. Therefore, the Winston-Lutz test should precede stereotactic radiosurgery for isocenter verification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.