Abstract

HyperArc is a preferred technique for treating brain metastases, employing a single isocenter for multiple lesions. Geometrical isocentricity in the TrueBeam linear accelerator with HyperArc is crucial. We evaluated machine performance checks (MPCs) as an alternative to the Winston-Lutz (WL) test to verify the treatment isocenter. Between January and July 2023, we assessed 53 data points using MPC and Winston-Lutz tests. The isocenter size obtained from the MPC and its sum, including the rotation-induced couch shift, were compared with the maximum total delta value from the Winston-Lutz test. The maximum total delta was 0.68 ± 0.10 mm, while the isocenter size was 0.28 ± 0.02 mm. The sum of the isocenter size and rotation-induced couch shift measured by MPC was 0.61 ± 0.03 mm. During the Winston-Lutz test (without couch rotation), the maximum total delta value was 0.56 ± 0.13 mm. A t-test analysis revealed a significant difference in the isocenter size averages between the Winston-Lutz and MPC outcomes, whereas the Pearson's correlation coefficient yielded no correlation. Our study highlights the necessity for separate MPC and Winston-Lutz tests for isocenter verification. Therefore, the Winston-Lutz test should precede stereotactic radiosurgery for isocenter verification.