Abstract

This study proposed an off-line method to quantitively analyze the dosimetric impact of daily setup variations on Intensity-modulated radiotherapy (IMRT) when these variations are not corrected due to the lack of image guidance. Series of daily Cone-beam computed tomography (CBCT) images and treatment plan data of a patient with esophageal cancer previously treated using Image-guided radiotherapy (IGRT) were analyzed retrospectively. Daily setup shifts were applied to IMRT plans on a fraction-by-fraction basis using six-degree-of-freedom rigid image registration. The mean interfraction translation displacement was 2.3 ± 1.8, 2.1 ± 0.9, and -1.8 ± 2.6 mm in the LR, AP, and SI directions, respectively. The average interfraction rotation displacement was 2.2 ± 1.2, -0.7 ± 0.8, and -0.7 ± 0.6 degree around the three orthogonal axes. Due to variations, a loss of equivalent uniform dose (EUD) up to 6% for planning target volume (PTV) was noted. Findings confirmed that successful implementation of high precision IMRT required daily image guidance to ensure accurate delivery of sharp gradient dose distributions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.