Abstract

A prototype laser-induced breakdown spectroscopy (LIBS) sensor is tested for the determination of rare earth elements (Eu and Yb) in liquid and solid samples. The sensor head, built using a monolithic passively Q-switched (PQSW) Nd:YAG laser, produced a 1064 nm laser beam with ns pulses and an energy of 4.2 mJ. The measurements show good calibration linearity for both Eu and Yb with R2 values above 0.99 for all analyzed spectral lines in liquid and solid samples. Limits of detection (LODs) obtained were as low as 1 ppm, which are comparable to or better than those reported previously by using table top actively Q-switched systems. This study aims to develop a high sensitivity, field deployable sensor for characterizing existing and new sources of rare earth elements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.