Abstract

Through numerical experimentation we investigated the isovelocity surface area flow convergence method used in estimating regurgitant valve flow rates. Recent advances in three-dimensional color Doppler flow imaging have created renewed interest in this method. Experimentation was based on the use of depth-averaged finite element models of the left heart. The heart models studied varied from “synthetic” representations to a model of a left heart traced from an actual echocardiographic image of a patient with a prolapsed mitral valve. The isovelocity surface area flow convergence method overestimated regurgitant flow rates throughout the Nyquist limits considered with a critical Nyquist limit in which this overestimation is minimized. The angle dependence of Doppler color flow imaging partially corrects for this overestimation. The isovelocity surface area flow convergence method is a viable alternative to methods currently in use. Through numerical experimentation, we have begun to shed light on the inaccuracies inherent in this flow convergence method. (J Am Soc Echocardiogr 1998;11:809-18.)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.