Abstract

Standard pulse oximeter auditory tones do not clearly indicate departures from the target range of oxygen saturation (SpO2) of 90%–95% in preterm neonates. We tested whether acoustically enhanced tones would improve participants' ability to identify SpO2 range. Twenty-one clinicians and 23 non-clinicians used (1) standard pulse oximetry variable-pitch tones plus alarms; (2) beacon-enhanced tones without alarms in which reference tones were inserted before standard pulse tones when SpO2 was outside target range; and (3) tremolo-enhanced tones without alarms in which pulse tones were modified with tremolo when SpO2 was outside target range. For clinicians, range identification accuracies (mean (SD)) in the standard, beacon, and tremolo conditions were 52% (16%), 73% (14%) and 76% (13%) respectively, and for non-clinicians 49% (16%), 76% (13%) and 72% (14%) respectively, with enhanced conditions always significantly more accurate than standard. Acoustic enhancements to pulse oximetry clearly indicate departures from preterm neonates’ target SpO2 range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.