Abstract
A multiple data point version of the industry standard, two data point raster-changing procedure is employed to measure low levels (< 1 x 1017 atoms/cm3) of nitrogen (N) in silicon carbide (SiC) by SIMS (Secondary Ion Mass Spectrometry). A current-changing procedure is also employed. Together, these are used evaluate the assumptions of the standard method, to separate and measure the components of background signal, and to improve upon the precision and accuracy of the standard method. The risk of poor precision in the two-point method is demonstrated, as is the improvement provided by the multiple-point method. Results show that, in addition to the wellknown N memory background, adsorption background can contribute significantly to the N signal. In general, establishing the presence of adsorption gas in this way can be used to warn of the presence of ionization background, which is not measurable per se.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.