Abstract

The Ag/AgCl reference electrode is commonly used in choline chloride based deep eutectic solvents. However, we found it undergoes significant potential shifts in electrochemical tests which previous reports largely ignored. In this work, we studied the degradation mechanism leading to its instability. Results show that due to the high Cl− concentration in ethaline, the AgCl film easily dissolves and forms AgCl2− species causing a potential shift. Therefore, we suggest a [Fe(CN)6]3−/[Fe(CN)6]4− reference electrode based on the reversibility and low diffusivity of [Fe(CN)6]3−/[Fe(CN)6]4− redox couple in ethaline, which was demonstrated to be reliable and stable over weeks of operation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.