Abstract

This study aims to determine adequacy of thick protective lead aprons as radiation in Radiology Services Unit in Semarang. This research is quantitative with observational approach radiation measurements. The samples were hospitals in Kota Semarang. The independent variable is the thickness of the lead apron and the dependent variable in this study is the radiation behind the lead apron. Data analysis was performed using descriptive statistics to take a conclusion. The results showed Adequacy lead apron at the Semarang Hospital, HVL values between 4 to 7 HVL or dose rate are forwarded after passing through the apron of 6.25% to 0, 781%. Value adequacy thick lead apron 0.5 mmPb that used had a value below the threshold set by Bapeten that dose forwarded after passing the lead apron is less than or <0,02 μSv which means lead apron can be used for radiation safety assurance in hospital and radiology departement units in the Semarang.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.