Abstract
The NucliSens Extractor in combination with the 2·0 version of the Roche Cobas HCV Amplicor test has been validated by five European blood screening laboratories in a multi-centre study. For testing the performance characteristics of this HCV-NAT method, the European Pharmacopoeia validation guidelines were followed. The CLB VQC reference reagents were used for testing robustness and sensitivity. After a technical improvement in the extraction stations, the NucliSens Extractor appeared to be contamination-free as was proved by testing negative controls alternating with samples containing a high HCV-RNA concentration. The Pelicheck HCV-RNA genotype 1 dilution panel was tested 74 times in the five laboratories and an overall 95% detection limit of 80 genome equivalents (geq)/ml was found. In one laboratory the Pelicheck panel was tested in 25 runs and here a 95% detection limit of 32geq/ml was achieved. In this laboratory the Pelispy HCV-RNA run control samples of 140geq/ml were consistently picked up in all extractor stations. In addition the laboratories have tested a WHO HCV-RNA genotype 1 standard dilution series 39 times and a Pelicheck HCV-RNA genotype 3 reference panel in 32 test runs. The limiting dilution analysis enabled us to compare the detection efficiency of the NucliSens-Amplicor method for the genoype 1 and genotype 3 isolates and to calibrate the reference reagents against each other. The combined Nuclisens-Amplicor method was found to detect the genotype 3 isolate in the Pelicheck HCV-RNA panels with 2–3 fold lower efficiency than the genotype 1 standard (assuming that the historical calibration of the genotype 3 against the genotype 1 standard is correct). In this study of a single method 1IU of the WHO HCV-RNA standard was found to be equivalent to 5·1geq of the VQC HCV-RNA standard (95% confidence intervals 3·1–9·1geq). To avoid confusion with the use of the CLB VQC reagents we accept the NIBSC collaborative study in which calibration by a variety of methods showed that the Pelispy 380geq/ml run control is equivalent to 100IU/ml of the WHO standard. This multi-centre validation study demonstrates that the 95% detection limit of the NucliSens HCV Amplicor method lies far below the detection limits required by the international regulatory bodies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.