Abstract

International sensitivity index (ISI) calibration of whole blood prothrombin time (PT) monitors is too complex. We previously simplified the method by using European Concerted Action on Anticoagulation (ECAA) lyophilized plasma samples with the TAS PT-NC (Bayer AG, Leverkusen, Germany) and the CoaguChek Mini (Roche Diagnostics, Mannheim, Germany) whole blood PT monitoring systems. The TAS PT-NC required a correction derived from the line of equivalence. Monte Carlo bootstrap analysis of reducing numbers of test samples was performed with both systems. Plasma samples from patients receiving coumarin (coumarin samples), healthy subjects (normal samples), and plasma samples artificially depleted of coagulation factors were used. With the TAS PT-NC, 20 coumarin samples or 20 artificially depleted samples with 7 normal samples gave reliable ISI and international normalized ratio and satisfactory precision. With the CoaguChek Mini, 30 coumarin and 10 normal samples were required. Simplification of ISI calibration of the 2 monitoring systems is possible using fewer ECAA lyophilized plasma samples than the 80 required according to the World Health Organization guidelines for conventional PT systems and previously recommended for fresh plasma samples tested on the same 2 monitoring systems.