Abstract

Some years ago the EURAMET.EM-S24 comparison was performed in the field of small DC currents below 1 nA, initiated and coordinated by PTB, to compare the capabilities of thirteen European and non-European National Metrology Institutes for traceable calibrations of picoammeters. Late in the comparison, SP announced an interest to participate; instead, the bilateral comparison EURAMET.EM-S38 between SP and PTB was started. The measurements were made before any report of the EM-S24 was published. Hence, the same travelling standards, two different commercial picoammeters, could be used. They were calibrated at current values of ±100 fA, ±1 pA, ±10 pA and ±100 pA. Good agreement between the results of the participants was found.Main text.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).