Abstract

The European Radiation Dosimetry Group (EURADOS) has been organising regular, self-funding, external dosimetry intercomparisons (ICs) since 2008. IC2019exteye, for extremity and eye lens dosemeters, attracted 113 systems (68 extremity and 45 eye lens) from 60 institutes with participants from 26 countries around the world. Irradiations were carried out in accredited European irradiation facilities in terms of Hp(0.07)/or Hp(3) in the ranges: photon energy: 16-662keV, beta mean energy 250-1000keV, dose: 0.5mSv to 1Sv, maximum angle of incidence: ±60°. This paper describes the IC2019exteye set-up, irradiation plan and coordination with an overview of the main conclusions. Out of the total of 113 systems, 57 reported results for photons only and 56 reported both photon and beta. In general, the participants showed very satisfactory performances with the medians of all Hp(0.07) and Hp(3) response values very close to unity.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call