Abstract

We report the bench testing of three new anaesthesia ventilators : Flexima™ (Datex), SA2/RA2™ and Cato™ (Dräger). The test circuit included a two compartment lung model, a pneumotachograph and a pressure gauge. Volume was integrated from flow signal. The tidal volume delivered by the Flexima™ and SA2/RA2™ ventilators decreased with increasing mechanical work load contrarily to the Cato™. Moreover, the tidal volume of the Flexima™ increased with the fresh gas flow. In spontaneous ventilation inspiratory resistance were low.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.