Abstract
Aqueous ethylene glycol (ethane 1,2 diol) as a green mobile phase has been used for thin layer chromatographic (TLC) studies of cationic surfactants on alumina layers. Nineteen solvent systems were used to examine the mobility of the surfactants and to discover the best TLC system for the selective separation of dodecyl trimethylammonium bromide (DTAB) from multi-component mixture of other surfactants. Among the TLC systems studied, M3 (ethylene glycol: water, 8:2) was best for achieving the selective separation of DTAB from multi-component mixture of other surfactants because in this mobile phase mobility of all surfactants except DTAB were insignificant. Effect of organic additives in aqueous ethylene glycol mobile phase on the mobility of surfactants was examined. The results obtained on laboratory made alumina TLC plates and commercially available precoated alumina HPTLC plates were compared. The lower limits of detection of DTAB, CPC, CTAB, HDTAC, and TTAB were 0.02, 0.05, 0.04, 0.06, or 0.08 µg per zone respectively. The resolution of mixture of cationic surfactants was also examined in the presence metal cations as an impurity in the analyzed sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.