Abstract

Background and objectiveThe corneal biomechanical properties in naïve, untreated glaucoma and ocular hypertension (OHT) eyes is interesting, because it may be a source of error in intraocular pressure (IOP) measurements by Goldmann applanation tonometer (GAT) and ocular response analyzer (ORA).The main objective of this study was to evaluate the IOP values obtained using GAT and the ORA, in primary open angle glaucoma (POAG) and in OHT untreated eyes. Material and methodsObservational, masked, cross sectional observational study. Newly diagnosed, untreated POAG and OHT eyes were included. ResultsFifty-one POAG and 34 OHT eyes were analyzed. We found that IOPcc (IOP corneal-compensated) was significantly higher than GAT IOP in POAG (P=.0002) while we did not find any significant difference between both tonometers in OHT (P=.1). ConclusionsGAT seems to underestimate the real IOP in untreated POAG eyes and it seems to be quite accurate in OHT eyes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.