Abstract

A problem with video signal analysis for estimating frequency-dependent ultrasonic attenuation is that relative echogenicity versus depth curves are distorted when broadband pulses are used. In this correspondence, we present results that demonstrate improved accuracy of attenuation estimates computed from B-mode or envelope data derived after narrowband (1 MHz BW) filtering at different frequencies around the center frequency of the broadband echo signal. Based on the premise that transducer center frequencies are selected in part on penetration or imaging depth requirements, simulation and experimental results for a typical ultrasound imaging system show that narrowband video signal analysis for frequencies lower than or at the center frequency of the broadband pulse provide unbiased attenuation estimation over this depth. Filtered signals above the center frequency of the pulse yield accurate results only at shallow depths.