Abstract

Strain rate (SR) measurement of heart tissue based on ultrasound images provides useful information of tissue hardness for diagnosing heart diseases. However, current SR estimation methods utilizing speckle tracking technique are based on conventional delay and sum (DS) imaging method, which causes skewed heart image resulting in inaccurate SR. To overcome the problem, a method to combine high frame rate (HFR) imaging method with speckle tracking technique was proposed. Using only one or a few transmissions for each image; compared with 91 for DS, new method can get a snapshot of moving targets, avoiding the skewing problem in DS. Two studies, with simulated and experimental echo data, respectively, were performed to verify the method. Both plane wave and limited diffraction beam (LDB) were studied for HFR. Both SR estimations in lateral and axial directions were calculated for the new and conventional ultrasound imaging methods. Results show that the new method has comparable or lower velocity errors than DS and more accurate SR, especially in lateral direction. Moreover, it can measure high velocity for other applications such as blood flow measurement. With the full view of heart image, SR of interest can be localized and then accurately estimated.