Abstract
Two methods can be used to assess the intra-cuff pressure of tracheostomy tubes: digital palpation of the pilot balloon and use of a hand-held manometer. We conducted a telephone survey to determine the prevalence of both methods in intensive care units within 21 teaching hospitals across the United Kingdom. Forty-two per cent of the intensive care units surveyed used a protocol for monitoring cuff pressure with a manometer.A study to compare these two methods, using the manometer as the reference standard, was then carried out. The cuff pressure was correctly estimated in pre-inflated tracheostomy tubes, in a tracheal model, by 61 per cent of a cross-section of intensive care unit and otolaryngology staff.Using pilot balloon palpation is inaccurate and leaves a significant proportion of patients at risk of tracheal injury. We advocate the wider availability of hand-held pressure manometers in intensive care units and the institution of protocols for monitoring cuff pressure for any patient with a tracheostomy tube with an inflated cuff in situ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.