Abstract

Time domain correlation technique is a widely used method for blood flow velocity measurement. The time shift between a pair of windowed ultrasonic echoes is estimated by searching the temporal position of the maximum of the interpolated normalized correlation function. Between two consecutive echoes, the acoustical footprint of a group of scatterers, which are transported with the flow, moves and is deformed. This implies a decreasing of the amplitude of the normalized correlation coefficient. In the case of microcirculation (low flow rate, low SNR), the amplitude of the correlation peak can be used to detect the presence of blood flow and to discriminate false and true detections (reliability index). We have numerically evaluated the statistical performances of the cross-correlation algorithm used as a correlation peak amplitude estimator in severe conditions (short correlation window length, low SNR). These theoretical results have been compared with in vitro experimentation on a 100-/spl mu/m-diameter microcirculatory phantom and with in vivo experimentation on a 180-/spl mu/m-diameter vessel of a human leg carrying erysipelas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.