Abstract

At the Neutron Calibration Laboratory (LCN) of IPEN/CNEN, a 241AmBe source was used to test and calibrate neutron detectors. The neutrons emitted by the source reach the detector as intended, but they also scatter undesirably from the building's floor, ceiling, and walls, leading to indirect detection. A Long Counter (LC) detector was modeled using the MCNPX Monte Carlo code. The objective of this study was to measure the contribution of scattered radiation at the LCN / IPEN, and to determine the fluence rate, at different points in the calibration room at source-to-detector distances of 100 cm and 150 cm; subsequently, the results were compared with those of the Brazilian Laboratory of Metrology of Ionizing Radiation (LNMRI). The results show that the fluence rates of LCN / IPEN are comparable to those of this neutron laboratory for the 100 and 150 cm source-to-detector distances using a 241AmBe source (37 GBq). The results show that the calibration of neutron detectors should be performed at distances less than 100 cm, where the contribution of scattered radiation is within the 40% limit recommended by ISO 8529-1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.