Abstract

To evaluate the possibility of respiratory-volume measurement using photoplethysmography (PPG), PPG signals from 16 normal volunteers are collected, and the respiratory-induced intensity variations (RIIV) are digitally extracted. The RIIV signals are studied while respiratory volume is varied. Furthermore, respiratory rate, body posture and type of respiration are varied. A Fleisch pneumotachograph is used as the inspired volume reference. The RIIV and pneumotachograph signals are compared, and a statistical analysis is performed (linear regression and t-tests). The key idea is that the amplitude of the RIIV signal is related to the respiratory volume. The conclusion from the measurements is that there exists a relationship between the amplitude of the RIIV signal and the respiratory volume (R = 0.842, s = 0.428, p < 0.005). Absolute measurements of the respiratory volume are not possible from the RIIV signal with the present set-up. The RIIV signal also seems to be affected by respiratory rate and type. More knowledge about respiratory parameters and improved sensor and filter design are required to make absolute measurements of volumes possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.