Abstract
Limit of detection (LOD) is usually determined by calculation of the standard deviation of measurements of blank samples. In several technical reports, LOD is defined as the concentration at the relative standard deviation (RSD) 30% (or other values) for repeated measurements. Similarly, limit of quantitation (LOQ) is also defined at the RSD level 10%. When a laboratory starts to use a method, the method verification should be carried out in advance to guarantee this method can be implemented correctly and the method characteristics (eg. LOD and LOQ) can be satisfied in this lab. But in routine test, this verification is not conducted every time and the measurement result might be effected by several factors. Then the verification results can vary under different test conditions. In this article the critical value of relative standard deviation is presented, when the RSD from routine test of blank samples of n times larger than the critical value it can be concluded that this measurement system cannot meet the requirement of LOD/LOQ that defined by the methods related with RSD. This method shows a quick check whether the routine measurement can satisfy the defined LOD/LOQ.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.