Abstract
The purpose of the paper is to report on the dose estimation studies at various depths for the commonly used beta-emitting 90Y, 166Ho, 153Sm, 32p, and 177Lu radionuclides in a phantom fabricated using the poly methyl methacrylate tissue equivalent material having a density of 1.19 gm/cc, by using thermoluminescent dosemeters. GAFChromic MD-55 films were used to calculate the calibration factor for the thermoluminescent micro-rods used in our study. It is observed that 90Y delivers the highest dose at 1 mm amongst the radionuclides tested followed by 32p, 166Ho, 153Sm, and 177Lu, whereas the cumulative dose received by the joint was found to be more for 32P followed by 90Y, 166Ho, 153Sm, and 177Lu. The highest therapeutic range obtained is 3.1 mm for 153Sm amongst the tested radionuclides. The dose values obtained for all the above-mentioned radionuclides can serve as reference material for those researchers and clinicians who are interested in selection of the radionuclide for the type of joint treated and the amount of dose necessary to be delivered to the synovial membrane.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.