Abstract

BackgroundPlasma HIV-1 RNA levels (pVLs), routinely used for clinical management, are influenced by measurement error (ME) due to physiologic and assay variation.ObjectiveTo assess the ME of the COBAS HIV-1 Ampliprep AMPLICOR MONITOR ultrasensitive assay version 1.5 and the COBAS Ampliprep Taqman HIV-1 assay versions 1.0 and 2.0 close to their lower limit of detection. Secondly to examine whether there was any evidence that pVL measurements closest to the lower limit of quantification, where clinical decisions are made, were susceptible to a higher degree of random noise than the remaining range.MethodsWe analysed longitudinal pVL of treatment-naïve patients from British Columbia, Canada, during their first six months on treatment, for time periods when each assay was uniquely available: Period 1 (Amplicor): 08/03/2000–01/02/2008; Period 2 (Taqman v1.0): 07/01/2010–07/03/2012; Period 3 (Taqman v2.0): 08/03/2012–30/06/2014. ME was estimated via generalized additive mixed effects models, adjusting for several clinical and demographic variables and follow-up time.ResultsThe ME associated with each assay was approximately 0.5 log10 copies/mL. The number of pVL measurements, at a given pVL value, was not randomly distributed; values ≤250 copies/mL were strongly systematically overrepresented in all assays, with the prevalence decreasing monotonically as the pVL increased. Model residuals for pVL ≤250 copies/mL were approximately three times higher than that for the higher range, and pVL measurements in this range could not be modelled effectively due to considerable random noise of the data.ConclusionsAlthough the ME was stable across assays, there is substantial increase in random noise in measuring pVL close to the lower level of detection. These findings have important clinical significance, especially in the range where key clinical decisions are made. Thus, pVL values ≤250 copies/mL should not be taken as the “truth” and repeat pVL measurement is encouraged to confirm viral suppression.

Highlights

  • Plasma Human immunodeficiency virus of type1 (HIV-1) RNA levels, routinely used for clinical management, are influenced by measurement error (ME) due to physiologic and assay variation

  • Model residuals for plasma HIV-1 RNA level (pVL) 250 copies/mL were approximately three times higher than that for the higher range, and pVL measurements in this range could not be modelled effectively due to considerable random noise of the data

  • The funding sources had no role in the choice of methods, the contents or form of this work, or the decision to submit the results for publication

Read more

Summary

Objective

To assess the ME of the COBAS HIV-1 Ampliprep AMPLICOR MONITOR ultrasensitive assay version 1.5 and the COBAS Ampliprep Taqman HIV-1 assay versions 1.0 and 2.0 close to their lower limit of detection. To examine whether there was any evidence that pVL measurements closest to the lower limit of quantification, where clinical decisions are made, were susceptible to a higher degree of random noise than the remaining range

Methods
Results
Conclusions