Abstract

The aim was to obtain inhomogenity correction factors (ICFs) for lung tissue inhomogenity for a Co-60 teletherapy beam using Monte Carlo simulation and to compare them with factors obtained from a commercially available treatment planning system. The Monte Carlo simulation code of EGSnrc is used for the depth dose calculations. Two clinical like situations were simulated-dose calculation point beyond the lung tissue volume and dose calculation point within the lung tissue volume. The variation of ICF with lung thicknesses and positions was studied. ICF values were obtained for the similar situations from a commercially available treatment planning system, Theraplan Plus. Percentage depth dose data obtained from Monte Carlo simulation is well matching with the published measurement data. ICFs for lung tissue inhomogenity calculated using the Monte Carlo code are in good agreement with Theraplan Plus TPS values for small inhomogenity thicknesses. These results can be used for the verification of TPS calculation or manual treatment time calculation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.