Abstract

Phase transition temperatures can be used for thermometer calibrations in accordance with the ITS-90. For this the thermometer is inserted into a fixed-point cell filled with a highly pure substance. While the substance is melting or freezing the temperature inside the cell remains almost constant and the thermometer measures a plateau in the temperature curve. From this plateau the phase transition temperature needs to be estimated for the calibration of the thermometer. This can be done using different mathematical methods, taking into account various systematic deviations as well as reproducibilities of the results depending on the chosen method, the fixed-point material’s purity and above all the amount of fixed-point material. Hence this article presents results from measurements in miniature fixed-point cells filled with zinc of various purities. The plateau curves were measured at different heating rates and comparatively analysed using five different estimation methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.