Abstract

The issues of detection of reference points of the arterial blood pulsation signal under the influence of physical and physiological artifacts are considered. Comparison of schemes for detecting the reference point of the arterial blood pulsation signal is carried out. As a criterion for the effectiveness of the scheme for detecting reference points of the arterial blood pulsation signal, the methodological error in determining the temporary position of the reference point from the true value was chosen. It has been shown that among the parametric means for detecting the reference point of the arterial blood pulsation signal, the most effective is the amplitude-time detector based on synchronous processing of ECG signals and arterial blood pulsation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.