Abstract

Simple, specific, accurate and precise method, namely, reverse phase high performance liquid chromatography was developed for estimation of duloxetine HCl in pharmaceutical formulations. For the high performance liquid chromatography method, Phenomenox C-18, 5 μm column consisting of 250×4.6 mm i.d. in isocratic mode, with mobile phase containing 0.01M 5.5 pH phosphate buffer: acetonitrile (60:40 v/v) and final pH adjust to 5.5±0.02 with phosphoric acid was used. The flow rate was 1.2 ml/min and effluent was monitored at 231 nm. The retention time was 5.61 min. The method was validated in terms of linearity, accuracy and precision. The linearity curve was found to be linear over 0.25-4 μg/ml. The limit of detection and limit of quantification were found to be 0.10 and 0.25 μg/ml respectively. The proposed method was successfully used to determine the drug content of marketed formulations.

Highlights

  • Specific, accurate and precise method, namely, reverse phase high performance liquid chromatography was developed for estimation of duloxetine HCl in pharmaceutical formulations

  • For the high performance liquid chromatography method, Phenomenox C-18, 5 μm column consisting of 250×4.6 mm i.d. in isocratic mode, with mobile phase containing 0.01M 5.5 pH phosphate buffer: acetonitrile (60:40 v/v) and final pH adjust to 5.5±0.02 with phosphoric acid was used

  • A few methods in literature were reported for the determination of Duloxetine HCl (DLX) and its key intermediate, desmethyl duloxetine in human serum by HPLC method[4-5]

Read more

Summary

Introduction

Et al.: RP-HPLC estimation of duloxetine hydrochloride Specific, accurate and precise method, namely, reverse phase high performance liquid chromatography was developed for estimation of duloxetine HCl in pharmaceutical formulations. For the high performance liquid chromatography method, Phenomenox C-18, 5 μm column consisting of 250×4.6 mm i.d. in isocratic mode, with mobile phase containing 0.01M 5.5 pH phosphate buffer: acetonitrile (60:40 v/v) and final pH adjust to 5.5±0.02 with phosphoric acid was used.

Results
Conclusion