Abstract

To improve estimation of cochlear implant (CI) insertion depth in postoperative skull x-rays using synthesized information from preoperative CT scans. Retrospective cohort. Tertiary referral center. Ten adult cochlear implant recipients with preoperative and postoperative temporal bone computed tomography (CT)scans and postoperative skull x-ray imaging. Postoperative x-rays and digitally reconstructed radiographs (DRR) from preoperative CTs were registered using 3D Slicer and MATLAB to enhance localization of the round window and modiolus. Angular insertion depth (AID) was estimated in unmodified and registration-enhanced x-rays and DRRs in the cochlear view. Linear insertion depth (LID) was estimated in registered images by two methods that localized the proximal CI electrode or segmented the cochlea. Ground truth assessments were made in postoperative CTs. Errors of insertion depth estimates were calculated relative to ground truth measurements and compared with paired t t ests. Pearson correlation coefficient was used to assess inter-rater reliability of two reviewer's measurements of AID in unmodified x-rays. In postoperative x-rays, AID estimation errors were similar with and without registration enhancement (-1.3 ± 20.7° and -4.8 ± 24.9°, respectively; mean ± SD; p = 0.6). AID estimation in unmodified x-rays demonstrated strong interrater agreement (ρ = 0.79, p < 0.05) and interrater differences (-15.0 ± 35.3°) comparable to estimate errors. Registering images allowed measurement of AID in the cochlear view with estimation errors of 14.6 ± 30.6° and measurement of LID, with estimate errors that were similar between proximal electrode localization and cochlear segmentation methods (-0.9 ± 2.2 mm and -2.1 ± 2.7 mm, respectively; p = 0.3). 2D-3D image registration allows measurement of AID in the cochlear view and LID using postoperative x-rays and preoperative CT imaging. The use of this technique may reduce the need for postimplantation CT studies to assess these metrics of CI electrode position. Further work is needed to improve the accuracy of AID assessment in the postoperative x-ray view with registered images compared with established methods.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call