Abstract
Although arterial mechanical properties have been evaluated using arterial input impedance, the relative difficulty in accurate flow measurement made the value of impedance somewhat limited. To develop an alternative method to evaluate arterial mechanical properties, we analyzed the aortic pressure (AoP)-radial arterial pressure (RAP) relationship because of relative ease in obtaining peripheral pressure waveform by tonometry. In 8 patients we simultaneously recorded aortic root and radial arterial pressure waveforms. We calculated the transfer function from AoP to RAP. We then fitted the transfer function to a model of a lossless uniform elastic tube terminated with a modified Windkessel model. The fact that the model transfer function predicted AoP from RAP waveform with considerable accuracy indicated that the model represented the arterial mechanical properties well. We conclude that we can estimate the arterial mechanical properties by the pressure-pressure transfer function analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.