Abstract

Tracheal and arterial CO2 partial pressures were measured simultaneously in 27 laryngectomized patients both while they were awake and during high-frequency jet ventilation. Tracheal gas was sampled during brief interruptions of high-frequency jet ventilation. Agreement between tracheal and arterial CO2 partial pressures was assessed using the Bland-Altman method. The tracheal-arterial CO2 partial pressures gradient during spontaneous breathing was significantly lower (P<0.0002) than during high-frequency jet ventilation. During spontaneous ventilation, the bias was −0.77 kPa (95% CI= −0.99 to −0.55 kPa), and the upper and lower limits of agreement were 0.29 kPa (95% CI = −0.11 to −0.7 kPa) and −1.83 kPa (95% CI = −2.24 to −1.43 kPa). During high-frequency jet ventilation, the bias was −1.61 kPa (95% CI = −1.76 to −1.46 kPa), and the limits of agreement were −0.48 kPa (95% CI = −0.75 to −0.21 kPa) and −2.74 kPa (95% CI = −3.01 to −2.47 kPa). Despite the poor agreement between tracheal CO2 partial pressure and arterial CO2 partial pressure, it is sufficient to allow for adjustment of ventilator settings during jet ventilation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.