Abstract

The objective of this study is to establish an empirical formula of full-energy peak efficiency (FEPE), for a Broad Energy High Purity Germanium (BEGe) type detector. FLUKA Monte Carlo-based code is used in this work to simulate radiation transport in the BEGe detector, and FEPE values have been estimated for various gamma energies and wide range of commercially available detector configurations. Then, using machine learning-based non-linear regression models, an empirical formula relating FEPE values with detector configuration (active volume and diameter-to-length ratio) and incident gamma radiation (energy >122 keV) is obtained. The results show that the effect of diameter-to-length ratio on FEPE values cannot be ignored in the case of a BEGe detector, which is not observed in case of coaxial type detectors. At lower energies, the efficiency of the BEGe detector is higher than the coaxial detector due to its large diameter-to-length ratio. This empirical formula can be used to quickly estimate FEPE values and verify nominal geometric parameters provided by the manufacturer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.