Abstract

In a recent report Sarmiento et al. (1976) presented a method for estimating the statistical error associated with 222Rn scintillation counting. Because of certain approximations, the method is less accurate than that of an earlier work by Lucas and Woodward (1964). We have compared the Sarmiento method and the Lucas method, and determined the magnitude of errors incurred using the approximations. For counting times greater than 300 minutes, the disadvantage of the slight inaccuracies of the Sarmiento method are outweighed by the advantage of easier calculation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.