Abstract

This study provides Heart Rate (HR) Estimation using wrist-type Photoplethysmogpraphy (PPG) sensor while the subject is running. We propose the algorithm to estimate heart rate for the wrist-type PPG sensor. Since body motion artifacts easily affect the arm portion, our method in this study also uses accelerometer built in the wrist-type sensor to improve the accuracy of heart rate estimation. Our method has two components. One is rejecting artifacts with the power spectrum's difference between PPG and acceleration obtained by frequency analysis. The other is the reliability of heart rate estimation, defined by the acceleration. Experimental results while our test subjects were running came closer to the holter Electrocardiogram (ECG) in high accuracy (r = 0.98, SD = 8.7 bpm). We, therefore, report the heart rate estimation method which has a higher degree of usability compared to existing methods using ECG.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.