Abstract

The present paper discusses the establishment of traceability of reference grade hydrometers at National Physical Laboratory, India (NPLI). The reference grade hydrometers are calibrated and traceable to the primary solid density standard. The calibration has been done according to standard procedure based on Cuckow's Method and the reference grade hydrometers calibrated covers a wide range. The uncertainty of the reference grade hydrometers has been computed and corrections are also calculated for the scale readings, at which observations are taken.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.