Abstract

Conventional comparison techniques used between National Metrology Institutes are not practicable for short-lived radionuclides because of geographical separations and transport difficulties. The NPL Secondary Standard Radionuclide Calibrator provides an alternative approach and a comparison was conducted with 18F to investigate its feasibility. The exercise was successful and the paper details the protocol used, the quality assurance mechanisms introduced to underpin the comparison and an analysis of the results. It was also demonstrated that this approach could be linked to the BIPM SIR system. Recommendations are presented for the extension of this work to other suitable, short-lived radionuclides.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.