Abstract

A further-developed three-plate rotation method [Appl. Opt. 31, 3767 (1992)] for the absolute testing of flats is used as a measuring method for establishing flatness standards. A special phase-stepping Fizeau interferometer was built that permits flats of diameters up to 200 mm to be tested. First results show a mean error between 0.002 and 0.003 λ in the determination of the absolute flatness deviations (λ = 632.8 nm). These deviations are obtained at 500-600 points on each plate. A number of experimental conditions connected with the thermal and mechanical stability of the plates must be fulfilled.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.