Abstract

In the present paper we investigate the features of IRGANOX® 1076 phenols as a material for electron spin resonance (ESR) dosimetry. We experimentally analyzed the ESR response of pellets of IRGANOX® 1076 phenols irradiated with 60Co photons. The best experimental parameters (modulation amplitude and microwave power) for dosimetric applications have been obtained. The dependence of ESR signal as function of γ dose is found to be linear in the dose range studied (12–60Gy) and the lowest measurable dose is found to be of the order of 1Gy. The signal after irradiation is very stable in the first thirty days. From the point of view of the tissue equivalence, these materials have mass energy absorption coefficient values comparable with those of soft tissue.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.