Abstract

BackgroundWe designed a new Esophageal Magnetic Anastomosis Device (EMAD) for thoracoscopic repair of esophageal atresia (EA) with tracheoesophageal fistula (TEF) without the need of handheld suturing or additional gastrostomy. MethodsSynthetic EA-TEF model: Spherical and tubular shaped rubber balloons and a term infant sized plastic doll were used. Medical students (n = 10) and surgical trainees (n = 10) were asked to perform thoracoscopic repair of an “EA” with a hand sutured anastomosis (HA) and with the EMAD. Euthanized animal model: The esophagus in 5 piglets (3–4 kg) was dissected and a thoracoscopic esophageal magnetic anastomosis (EMA) was performed. Bursting pressure (BP) and pulling force (PF): HA and EMA were created on ex vivo New Zealand white rabbit (2.5–3 kg) esophagi (n = 25 in each test series). BP and PF were measured and compared against each other. ResultsMedical students were unable to complete HA, but were successful with the EMAD in 11.1 ± 2.78 min. Surgical trainees completed EMA in 4.6 ± 2.06 min vs. HA 30.8 ± 4.29 min (p<0.001). The BP following a HA (14.1 ± 3.32 cmH2O) was close to the physiological intraluminal pressure reported in a neonatal esophagus (around 20 cmH2O), whereas the BP with the EMAD was extremely high (>90 cmH2O) (p<0.001). The PF of an EMA (1.8 ± 0.30 N) was closer to the safety limits of anastomotic tension reported in the literature (i.e. 0.75 N) compared with the HA (3.6 ± 0.43 N) (p<0.0001). ConclusionThe EMAD could simplify, shorten, and potentially improve the outcome of thoracoscopic repair for EA with TEF in the future. A high BS and a relative low PF following EMAD application may lower the risk of postoperative complications such as esophageal leakage and stricture formation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call